The iMOST platform is manufactured in Shanghai, China, within ISO 13485–certified facilities. These sites have successfully undergone rigorous audits by global IVD partners.

No. Cloud connectivity is optional. All analysis is performed locally on the device.

CAP accredits laboratories, not individual devices. iMOST-U holds CE marking and SFDA marketing authorization (MDMA). For CAP- or CBAHI-accredited laboratories, iMOST-U can be integrated into their POCT programs under laboratory oversight, following applicable requirements for method verification, quality control, operator competency, and proficiency testing or alternative performance assessment.

The iMOST platform is not yet cleared or waived by the FDA. U.S. FDA 510(k) and CLIA waiver submissions are currently in preparation.

Future expansion includes the iMOST-U Dock and iMOST-U&C platforms, with planned tests covering:

* Hematology: CBC (19 parameters).
* Chemistry: CRP, Urine ACR, Diabetes Kidney Panel, Liver Function Panel, Lipid Panel, and Electrolytes.
* Immunology: HIV CD4.

Troubleshooting is typically initiated remotely. If hardware service is necessary, support is coordinated through the local distributor or an authorized service partner, adhering to the established service agreement.

N/A in certain parameter in “Results”

Possible Cause:

Certain parameters are rejected due to improper sample, improper card, or defects on the card.

Action Required:

Use a new Q-Card, repeat the measurement.